Standard Databases
Database Name | Sex | Protocol | Cases | Age | BMI | Acquisition parameters | Orientation | Isotope and dosage | Camera | Attenuation correction | Reconstruction parameters |
---|---|---|---|---|---|---|---|---|---|---|---|
FemaleProneStressMB | Female | Stress | 40 | 56 ± 14 | 27 ± 5 | non circular 180°, 64 projections, 15 seconds per projection |
Prone | 99mTc-Sestamibi, 25-40 mCi based on patient’s weight |
Philips Adac Forte, Philips Vertex, Siemens E-Cam |
N/A | Iterative Reconstruction (12 iterations), Butterworth Prefiltering (cutoff-0.55 cycle/pixel, order 5) |
FemaleRestMB-AC | Female | Rest | 50 | 55.9 ± 13.1 | 30.9 ± 6.3 | non circular 180°, 64 projections, 45-50 seconds per projection |
Supine | 99mTc-Sestamibi, 8.5-11.6 mCi based on patient’s weight |
Philips Vertex | Vantage Pro (hardware/software) with 2 Gadolinium-153 scanning line sources | AutoSPECT and Vantage Pro pre-processing |
FemaleRestMB | Female | Rest | 40 | 55.9 ± 13.1 | 30.9 ± 6.3 | non circular 180°, 64 projections, 45-50 seconds per projection |
Supine | 99mTc-Sestamibi, 8.5-11.6 mCi based on patient’s weight |
Philips Vertex | N/A | AutoSPECT and Vantage Pro pre-processing |
FemaleRestTL | Female | Rest | 40 | 56 ± 14 | 27 ± 5 | non circular 180°, 64 projections, 35 seconds per projection |
Supine | TL-201 Thallous Chloride, 3-3.5 mCi based on patient’s weight |
Philips Vertex, Philips Adac Forte, Siemens E-Cam |
N/A | Iterative Reconstruction (12 iterations), Butterworth Prefiltering (cutoff-0.66 cycle/pixel, order 5) |
FemaleRestTetro-AC | Female | Rest | 40 | 56.8 ± 9.3 | 29.9 ± 4.1 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 8-12 mCi based on patient weight |
GE Infinia Hawkeye | 4 slice CT | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
FemaleRestTetro | Female | Rest | 40 | 56.8 ± 9.3 | 29.9 ± 4.1 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 8-12 mCi based on patient weight |
GE Infinia Hawkeye | N/A | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
FemaleStressMB-AC | Female | Stress | 50 | 55.9 ± 13.1 | 30.9 ± 6.3 | non circular 180°, 64 projections, 30-40 seconds per projection |
Supine | 99mTc-Sestamibi, 29.5-42 mCi based on patient’s weight |
Philips Vertex | Vantage Pro (hardware/software) with 2 Gadolinium-153 scanning line sources | AutoSPECT and Vantage Pro pre-processing |
FemaleStressMB | Female | Stress | 40 | 56 ± 14 | 27 ± 5 | non circular 180°, 64 projections, 25 seconds per projection |
Supine | 99mTc-Sestamibi, 25-40 mCi based on patient’s weight |
Philips Vertex, Philips Adac Forte, Siemens E-Cam |
N/A | Iterative Reconstruction (12 iterations), Butterworth Prefiltering (cutoff-0.66 cycle/pixel, order 5) |
FemaleStressTL | Female | Stress | 40 | N/A | N/A | 55-60 seconds per projection | Supine | 4 mCi TL-201 Thallous Chloride | Picker Axis, Picker Prism |
N/A | FBP Reconstruction, Butterworth filtering (cutoff-0.66 cycle/pixel, order 10) |
FemaleStressTetro-AC | Female | Stress | 40 | 56.8 ± 9.3 | 29.9 ± 4.1 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 27-34 mCi based on patient weight |
GE Infinia Hawkeye | 4 slice CT | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
FemaleStressTetro | Female | Stress | 40 | 56.8 ± 9.3 | 29.9 ± 4.1 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 27-34 mCi based on patient weight |
GE Infinia Hawkeye | N/A | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
MaleProneStressMB | Male | Stress | 40 | 56 ± 14 | 27 ± 5 | non circular 180°, 64 projections, 15 seconds per projection |
Prone | 99mTc-Sestamibi, 25-40 mCi based on patient’s weight |
Philips Adac Forte, Philips Vertex, Siemens E-Cam |
N/A | Iterative Reconstruction (12 iterations), Butterworth Prefiltering (cutoff-0.55 cycle/pixel, order 5) |
MaleRestMB-AC | Male | Rest | 50 | 55.9 ± 13.1 | 30.9 ± 6.3 | non circular 180°, 64 projections, 45-50 seconds per projection |
Supine | 99mTc-Sestamibi, 8.5-11.6 mCi based on patient’s weight |
Philips Vertex | Vantage Pro (hardware/software) with 2 Gadolinium-153 Scanning line sources | AutoSPECT and Vantage Pro Pre-processing |
MaleRestMB | Male | Rest | 40 | 55.9 ± 13.1 | 30.9 ± 6.3 | non circular 180°, 64 projections, 45-50 seconds per projection |
Supine | 99mTc-Sestamibi, 8.5-11.6 mCi based on patient’s weight |
Philips Vertex | N/A | AutoSPECT and Vantage Pro Pre-processing |
MaleRestTL | Male | Rest | 40 | 56 ± 14 | 27 ± 5 | non circular 180°, 64 projections, 35 seconds per projection |
Supine | TL-201 Thallous Chloride, 3-3.5 mCi based on patient’s weight |
Philips Vertex, Philips Adac Forte, Siemens E-Cam |
N/A | Iterative Reconstruction (12 iterations), Butterworth Prefiltering (cutoff-0.66 cycle/pixel, order 5) |
MaleRestTetro-AC | Male | Rest | 40 | 49.7 ± 7.9 | 28.7 ± 4.7 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 8-12 mCi based on patient weight |
GE Infinia Hawkeye | 4 slice CT | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
MaleRestTetro | Male | Rest | 40 | 49.7 ± 7.9 | 28.7 ± 4.7 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 26-33 mCi based on patient weight |
GE Infinia Hawkeye | N/A | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
MaleStressMB-AC | Male | Stress | 50 | 55.9 ± 13.1 | 30.9 ± 6.3 | non circular 180°, 64 projections, 30-40 seconds per projection |
Supine | 99mTc-Sestamibi, 29.5-42 mCi based on patient’s weight |
Philips Vertex | Vantage Pro (hardware/software) with 2 Gadolinium-153 scanning line sources | AutoSPECT and Vantage Pro pre-processing |
MaleStressMB | Male | Stress | 40 | 56 ± 14 | 27 ± 5 | non circular 180°, 64 projections, 25 seconds per projection |
Supine | 99mTc-Sestamibi, 25-40 mCi based on patient’s weight |
Philips Vertex, Philips Adac Forte, Siemens E-Cam |
N/A | Iterative Reconstruction (12 iterations), Butterworth Prefiltering (cutoff-0.66 cycle/pixel, order 5) |
MaleStressTL | Male | Stress | 40 | N/A | N/A | 55-60 seconds per projection | Supine | 4 mCi TL-201 Thallous Chloride | Picker Axis, Picker Prism |
N/A | FBP Reconstruction, Butterworth filtering (cutoff-0.66 cycle/pixel, order 10) |
MaleStressTetro-AC | Male | Stress | 40 | 49.7 ± 7.9 | 28.7 ± 4.7 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 26-33 mCi based on patient weight |
GE Infinia Hawkeye | 4 slice CT | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
MaleStressTetro | Male | Stress | 40 | 49.7 ± 7.9 | 28.7 ± 4.7 | 60 projections, 25 seconds per projection |
Supine | 99mTc-Tetrofosmin, 26-33 mCi based on patient weight |
GE Infinia Hawkeye | N/A | Iterative 3D reconstruction, Butterworth Filter (cutoff 0.3, order 10) for ungated images, Butterworth filter (cutoff 0.28, order 10) for gated images |
PET Databases
Database Name | Sex | Protocol | Cases | Age | BMI | Acquisition parameters | Orientation | Isotope and dosage | Camera | Attenuation correction | Reconstruction parameters |
---|---|---|---|---|---|---|---|---|---|---|---|
MaleFemaleRestPETRb-V2 | Unisex (Males-21 and Females-21) | Rest | 42 | 61 ± 14 | 28 ± 8 | List mode acquisition, CT Transmission scan, 6-minute rest imaging acquisition |
Supine | Rb-82, 25-50 mCi |
Siemens Biograph-64 TruePoint PET/CT with the TrueV option | CT-based attenuation correction | Emission data were reconstructed with 2D Attenuation Weighted Ordered Subset Expectation Maximization (2 iterations and 21 subsets). CT-based attenuation, scatter, including prompt gamma, decay, and random corrections were applied to the reconstructed images |
MaleFemaleStressPETRb-V2 | Unisex (Males-21 and Females-21) | Stress | 42 | 61 ± 14 | 28 ± 8 | List mode acquisition, CT Transmission scan, 6-minute stress imaging acquisition |
Supine | Rb-82, 25-50 mCi |
Siemens Biograph-64 TruePoint PET/CT with the TrueV option | CT-based attenuation correction | Emission data were reconstructed with 2D Attenuation Weighted Ordered Subset Expectation Maximization (2 iterations and 21 subsets). CT-based attenuation, scatter, including prompt gamma, decay, and random corrections were applied to the reconstructed images |
NH3Rest | Any | Rest | 15 | n/a | n/a | n/a | Any | 13N-ammonia 10-20 mCi |
n/a | n/a | n/a |
NH3Stress | Any | Stress | 15 | n/a | n/a | n/a | Any | 13N-ammonia 10-20 mCi |
n/a | n/a | n/a |
GE Databases
Database Name | Sex | Protocol | Cases | Age | BMI | Acquisition parameters | Orientation | Isotope and dosage | Camera | Attenuation correction | Reconstruction parameters |
---|---|---|---|---|---|---|---|---|---|---|---|
GE530cFemaleProneStressMB | Female | Stress | 32 | 45.5 ± 10.9 | 28.0 ± 8.1 | 60 projections, 3-5 minutes per scan |
Prone | 99mTc-Sestamibi, 15-30 mCi based on patient’s weight |
GE Discovery NM 530c | N/A | N/A |
GE530cFemaleRestMB | Female | Rest | 32 | 61.4 ± 10.0 | 28.0 ± 5.3 | 60 projections, 3-5 minutes per scan |
Supine | 99mTc-Sestamibi, 5-10 mCi based on patient’s weight |
GE Discovery NM 530c | N/A | N/A |
GE530cFemaleStressMB | Female | Stress | 32 | 45.5 ± 10.9 | 28.0 ± 8.1 | 60 projections, 3-5 minutes per scan |
Supine | 99mTc-Sestamibi, 15-30 mCi based on patient’s weight |
GE Discovery NM 530c | N/A | N/A |
GE530cMaleProneStressMB | Male | Stress | 32 | 43.3 ± 4.7 | 28.4 ± 5.6 | 60 projections, 3-5 minutes per scan |
Prone | 99mTc-Sestamibi, 15-30 mCi based on patient’s weight |
GE Discovery NM 530c | N/A | N/A |
GE530cMaleRestMB | Male | Rest | 32 | 51.6 ± 5.5 | 28.6 ± 4.5 | 60 projections, 3-5 minutes per scan |
Supine | 99mTc-Sestamibi, 5-10 mCi based on patient’s weight |
GE Discovery NM 530c | N/A | N/A |
GE530cMaleStressMB | Male | Stress | 32 | 43.3 ± 4.7 | 28.4 ± 5.6 | 60 projections, 3-5 minutes per scan |
Supine | 99mTc-Sestamibi, 15-30 mCi based on patient’s weight |
GE Discovery NM 530c | N/A | N/A |
Philips Databases
Database Name | Sex | Protocol | Cases | Age | BMI | Acquisition parameters | Orientation | Isotope and dosage | Camera | Attenuation correction | Reconstruction parameters |
---|---|---|---|---|---|---|---|---|---|---|---|
AstonishFemaleRest-AC | Female | Rest | 40 | 53.3 ± 7.8 | 30.3 ± 6.1 | non circular 180°, 64 projections, 30 seconds per projection |
Supine | 99mTc-Sestamibi, 10 mCi |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources | Gd-153 scanning line sources | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images. Attenuation maps were reconstructed by a Bayesian prior approach using 12 iterations with a uniform initial estimate. |
AstonishFemaleRest | Female | Rest | 40 | 53.3 ± 7.8 | 30.3 ± 6.1 | non circular 180°, 64 projections, 30 seconds per projection |
Supine | 99mTc-Sestamibi, 10 mCi |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources. | N/A | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images. Attenuation maps were reconstructed by a Bayesian prior approach using 12 iterations with a uniform initial estimate. |
AstonishFemaleStress-AC | Female | Stress | 40 | 53.3 ± 7.8 | 30.3 ± 6.1 | non circular 180°, 64 projections, 20-25 seconds per projection |
Supine | 99mTc-Sestamibi, 25-35 mCi based on patient’s weight |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources. | Gd-153 scanning line sources | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images. Attenuation maps were reconstructed by a Bayesian prior approach using 12 iterations with a uniform initial estimate. |
AstonishFemaleStress | Female | Stress | 40 | 53.3 ± 7.8 | 30.3 ± 6.1 | non circular 180°, 64 projections, 20-25 seconds per projection |
Supine | 99mTc-Sestamibi, 25-35 mCi based on patient’s weight |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources. | N/A | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images |
AstonishMaleRest-AC | Male | Rest | 40 | 53.9 ± 8.7 | 28.0 ± 5.5 | non circular 180°, 64 projections, 30 seconds per projection |
Supine | 99mTc-Sestamibi, 10 mCi |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources | Gd-153 scanning line sources | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images. Attenuation maps were reconstructed by a Bayesian prior approach using 12 iterations with a uniform initial estimate. |
AstonishMaleRest | Male | Rest | 40 | 53.9 ± 8.7 | 28.0 ± 5.5 | non circular 180°, 64 projections, 30 seconds per projection |
Supine | 99mTc-Sestamibi, 10 mCi |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources | N/A | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images. Attenuation maps were reconstructed by a Bayesian prior approach using 12 iterations with a uniform initial estimate. |
AstonishMaleStress-AC | Male | Stress | 40 | 53.9 ± 8.7 | 28.0 ± 5.5 | non circular 180°, 64 projections, 20-25 seconds per projection |
Supine | 99mTc-Sestamibi, 25-35 mCi |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources | Gd-153 scanning line sources | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images. Attenuation maps were reconstructed by a Bayesian prior approach using 12 iterations with a uniform initial estimate. |
AstonishMaleStress | Male | Stress | 40 | 53.9 ± 8.7 | 28.0 ± 5.5 | non circular 180°, 64 projections, 20-25 seconds per projection |
Supine | 99mTc-Sestamibi, 25-35 mCi |
Philips CardioMD with LEHR collimators and Gd-153 scanning line sources | N/A | Astonish reconstruction (3D-OSEM), 4 iterations, 8 subsets and a matched filter cutoff of 1.0 was used for projection images. Attenuation maps were reconstructed by a Bayesian prior approach using 12 iterations with a uniform initial estimate. |
Siemens Databases
Database Name | Sex | Protocol | Cases | Age | BMI | Acquisition parameters | Orientation | Isotope and dosage | Camera | Attenuation correction | Reconstruction parameters |
---|---|---|---|---|---|---|---|---|---|---|---|
IQSPECTRestACFemale | Female | Rest | 22 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
IQSPECTRestACMale | Male | Rest | 19 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
IQSPECTRestNCFemale | Female | Rest | 27 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
IQSPECTRestNCMale | Male | Rest | 25 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
IQSPECTStressACFemale | Female | Stress | 26 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
IQSPECTStressACMale | Male | Stress | 20 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
IQSPECTStressNCFemale | Female | Stress | 27 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
IQSPECTStressNCMale | Male | Stress | 29 | n/a | n/a | n/a | Supine | 99mTc-Sestamibi | Siemens | n/a | n/a |
ccamFemaleRestTL | Female | Rest | 25 | n/a | n/a | non circular 90°, 64 projections |
Supine | TL-201Thallous Chloride, 3-3.5 mCi based on patient’s weight |
Siemens c.cam with LEAP collimators and Gd-153 scanning line sources | n/a | n/a |
ccamFemaleStressMB | Female | Stress | 27 | n/a | n/a | non circular 90°, 64 projections, 30 seconds per projection |
Supine | 99mTc-Sestamibi | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
ccamMaleRestTL | Male | Rest | 26 | n/a | n/a | non circular 90°, 64 projections |
Supine | TL-201Thallous Chloride, 3-3.5 mCi based on patient’s weight |
Siemens c.cam with LEAP collimators and Gd-153 scanning line sources | n/a | n/a |
ccamMaleStressMB | Male | Stress | 26 | n/a | n/a | non circular 90°, 64 projections, 30 seconds per projection |
Supine | 99mTc-Sestamibi | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearFemaleRestTc_AC | Female | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearFemaleRestTc_NC | Female | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearFemaleStressTc_AC | Female | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearFemaleStressTc_NC | Female | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearMaleRestTc_AC | Male | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearMaleRestTc_NC | Male | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearMaleStressTc_AC | Male | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
cclearMaleStressTc_NC | Male | Stress | 40 | n/a | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi, | Siemens c.cam with LEHR collimators and Gd-153 scanning line sources | n/a | n/a |
symbiaFemaleRestTc_AC | Female | Rest | 37 | 54.2 ± 11.0 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaFemaleRestTc_NC_F3D | Female | Rest | 37 | 54.2 ± 11.0 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaFemaleRestTc_NC_FBP | Female | Rest | 37 | 54.2 ± 11.0 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaFemaleStressTc_AC | Female | Stress | 37 | 54.2 ± 11.0 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaFemaleStressTc_NC_F3D | Female | Stress | 37 | 54.2 ± 11.0 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaFemaleStressTc_NC_FBP | Female | Stress | 37 | 54.2 ± 11.0 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaMaleRestTc_AC | Male | Rest | 37 | 52.5 ± 8.8 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaMaleRestTc_NC_F3D | Male | Rest | 37 | 52.5 ± 8.8 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaMaleRestTc_NC_FBP | Male | Rest | 37 | 52.5 ± 8.8 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaMaleStressTc_AC | Male | Stress | 37 | 52.5 ± 8.8 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaMaleStressTc_NC_F3D | Male | Stress | 37 | 52.5 ± 8.8 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
symbiaMaleStressTc_NC_FBP | Male | Stress | 37 | 52.5 ± 8.8 | n/a | non circular 90°, 64 projections |
Supine | 99mTc-Sestamibi | Siemens Symbia | n/a | n/a |
Spectrum Dynamics Databases
Database Name | Sex | Protocol | Cases | Age | BMI | Acquisition parameters | Orientation | Isotope and dosage | Camera | Attenuation correction | Reconstruction parameters |
---|---|---|---|---|---|---|---|---|---|---|---|
D-SPECT Female Rest MIBI combined | Female | Rest | 60 | 56 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 99mTc-Sestamibi, 8-9mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |
D-SPECT Female Rest TL combined | Female | Rest | 22 | 57 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 201Thallous Choloride, 2-2.5mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |
D-SPECT Female Stress MIBI combined | Female | Stress | 60 | 56 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 99mTc-Sestamibi, 36-40mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |
D-SPECT Female Stress TL combined | Female | Stress | 22 | 57 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 201Thallous Choloride, 2-2.5mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |
D-SPECT Male Rest MIBI combined | Male | Rest | 60 | 56 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 99mTc-Sestamibi, 8-9mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |
D-SPECT Male Rest TL combined | Male | Rest | 30 | 57 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 201Thallous Choloride, 2-2.5mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |
D-SPECT Male Stress MIBI combined | Male | Stress | 60 | 56 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 99mTc-Sestamibi, 36-40mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |
D-SPECT Male Stress TL combined | Male | Stress | 30 | 57 ± 13 | 28 ± 7 | no rotation;, 64 projections |
Upright and Supine | 201Thallous Choloride, 2-2.5mCi based on patient’s weight |
Spectrum Dynamics, D-SPECT |
n/a | n/a |